Open original FDA record ↗Recall record
- FDA identifier
- Z-1655-2026
- Company in FDA record
- ReCor Medical Inc.
- FDA classification date
- Mar 27, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to firm distributing a nonconforming product that supposed to be scrapped and not for human use.
Source facts
- Action
- On 01/28/2026, the firm informed customers verbally and sent via mail an "MEDICAL DEVICE RECALL" Letter that nonconforming products (cracking in the hub adjacent to the guidewire port) that may lead to fluid leaks and could trigger a pressure error. Customers are instructed to: 1. Do not use the affected catheters for any patient procedures. 2. Identify the catheters from the affected lot and return to Recor Clinical Specialist through RMA process. 3. Review and complete the Acknowledgement and Receipt Form and provide to the Recor Clinical Specialist. For questions or further assistance - contact Director of Regulatory & Quality Compliance at Lisa.luu@recormedical.com or telephone at 650-542-7700
- Initiated
- 2026-01-29
- Root cause
- Under Investigation by firm
- Distribution
- U.S. Nationwide distribution in the states of CA, CO, FL, KS, MO, NC, OH, TX, and VA.
- Recall number
- Z-1655-2026
- Recall status
- Open, Classified
- CDRH record ID
- 218970
- FDA classified
- 2026-03-27
- Recall event number
- 98365