Open original FDA record ↗Recall record
- FDA identifier
- Z-1656-2010
- Company in FDA record
- Beckman Coulter Inc.
- FDA classification date
- May 26, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Issue 1: The Diagnostic procedure, Simulate-CDNR Cycle, is not functioning properly. Impact: There is a potential for inadequate sample preparation of samples following the use of this Diagnostic function impacting WBC, Differential, Reticulocyte and/or NRBC results. Issue 2: Use of the Auto Prune function causes the deletion of Quality Control files. Impact: If the last run for a control lot
Source facts
- Action
- Beckman Coulter, Inc. issued an "Urgent: Product Corrective Action" notification dated March 10, 2010. Consignees were informed of the affected product and instructed on proper steps to take and requested to return the enclosed acknowledgement form. For further information, contact your local Beckman Coulter Representative or call 1-800-526-7694.
- Initiated
- 2010-03-02
- Root cause
- Software design
- Terminated
- 2012-05-23
- Distribution
- Nationwide Distribution (AZ, CA, CO, FL, GA, IA, IL, IN, KS, LA, MO, NC, NJ, NV, NY, OH, PA, SC, TN, TX, UT, WA, and WI) and Canada.
- Recall number
- Z-1656-2010
- Recall status
- Terminated
- CDRH record ID
- 90277
- FDA classified
- 2010-05-26
- Recall event number
- 55194