Philips Medical Systems North America Co. Phillips
FDA classification date
Mar 16, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification
Reason stated in the source
A wire in the connector of the Generator Grid Switch Supervisor box might short-circuit between 12V power and ground, and it will cause a System breakdown.
Source facts
Action
On 10/15/2010, Philips began sending out the URGENT-Device Correction Notice Failure Causing the Generator to Stop Working notification letter to their consignees.
The consignees were informed that a short circuit may occur due to a faulty wire connection in the Supervisor Box Connector in the Generator, and it will cause the generator stops working. When the system fails during a critical interventional case, this may cause risk to patients (no Fluoro available).
The systems involved in this recall are Allura Xper with the Velara Generator, Allura Integris with OMCP Generator, MultiDiagnost Eleva, and OmniDiagnost Eleva Imaging system.
Consignees can contact Philips Healthcare Call Center at 800-722-9377, #5 and Reference FCO 70800114, 70800115, 72200184, and 72200185.
Initiated
2010-10-07
Root cause
Nonconforming Material/Component
Terminated
2012-02-23
Distribution
Devices were distributed through out the US.
Recall number
Z-1656-2011
Recall status
Terminated
CDRH record ID
97393
FDA classified
2011-03-16
Recall event number
57777