Open original FDA record ↗Recall record
- FDA identifier
- Z-1657-2008
- Company in FDA record
- ArthroCare Corporation
- FDA classification date
- Sep 16, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Devices for which sterility may be compromised as evidenced by punctures in the plastic tray during shipping. The risk to the patient is a potential infection subsequent to the surgical procedure.
Source facts
- Action
- Urgent Product Recall letters (Dated January 31, 2008) were issued to end users of the device (medical professionals) and direct distributors. The letter instructs users to return any affected products.
- Initiated
- 2008-01-31
- Root cause
- Packaging process control
- Terminated
- 2011-01-08
- Distribution
- Worldwide Distribution --- USA including states of AL, AR, AZ, CA, CO FL, GA, HI, IL, IN, KS, KY, LA, MA, MD, ME, MN, MO, NC, ND, NJ, NV, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA, WI, and countries of Sweden, Australia, Germany, Italy, UK, Austria, Saudi Arabia, Belgium, Portugal, UAE, Kuwait, Singapore, Ireland, Spain, Finland, South Korea, Russia and Vietnam.
- Recall number
- Z-1657-2008
- Recall status
- Terminated
- CDRH record ID
- 68631
- FDA classified
- 2008-09-16
- Recall event number
- 46962