Reason stated in the source
The glenoid implant may not seat fully on the face of the natural glenoid due to inadequate measurement tolerances (tolerance stack-up) in the design of the cutter. Use of the instrument may lead to incomplete seating of the glenoid implant, elevating the theoretical risk of glenoid loosening, which may lead to early revision surgery. The problem may contribute to early device failure and patien
Source facts
- Action
- Consignees were notified by Urgent Medical Device Removal Notice dated 3/13/08 to discontinue use of the instruments and return to the firm. Operating room managers were sent a separate letter dated 3/13/08 and informed of the possible effects to patients from previous use of this instrument and it was suggested they monitor patients for these effects. If you have any questions contact Mary Hardesty at 1-800-348-9500 or (574) 372-3983.
- Initiated
- 2008-03-13
- Root cause
- Device Design
- Terminated
- 2008-10-15
- Distribution
- Class II Recall - Nationwide Distribution.
- Recall number
- Z-1658-2008
- Recall status
- Terminated
- CDRH record ID
- 69084
- FDA classified
- 2008-09-16
- Recall event number
- 47238