Open original FDA record ↗Recall record
- FDA identifier
- Z-1658-2009
- Company in FDA record
- NRT - Nordisk Rontgen Teknik A/S
- FDA classification date
- Sep 29, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The GE Precision MPi is in violation with 21 CFR Part 1020.32 requiring a signal audible to the fluoroscopist to sound for each passage of 5 minutes of fluoroscopic irradiation time during an examination or procedure.
Source facts
- Action
- An Advisory Notice concerning GE Precision MPi, dated May 1, 2009, was issued to Healthcare Professionals. The notice discussed the problem, risk, recommended immediate action, future actio, and affected system and serial numbers. A GE Field Service Representative will visit each affected U.S. customer site and introduce a speaker connected to the generator touch screen (to which the audible tone is delivered) in the operator's room, to be placed on the wall in the x-ray room. For further information, customers were to contact their local GE field engineer.
- Initiated
- 2009-05-01
- Root cause
- Radiation Control for Health and Safety Act
- Terminated
- 2009-10-20
- Distribution
- Nationwide Distribution, including CA, CO, FL, IN, MA, MI, MN, NV, NY, OH, OR, PA, TN, VT, WA, WI.
- Recall number
- Z-1658-2009
- Recall status
- Terminated
- CDRH record ID
- 82399
- FDA classified
- 2009-09-29
- Recall event number
- 52173