Open original FDA record ↗Recall record
- FDA identifier
- Z-1658-2010
- Company in FDA record
- Beckman Coulter Inc.
- FDA classification date
- May 26, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Issue 1: The system will substitute or omit the characters # @ 1 \ ] ' ( I ) - when the instrument is configured in a language other than English, Chinese or Japanese. Example: Using the character @, a substitution in the string '123@456' would result in '123#456' and an omission would result in '123456'. Issue2: The characters 7 " and * (question mark, quotation mark, and asterisk) used as part
Source facts
- Action
- Beckman Coulter, Inc. issued "Urgent: Product Corrective Action" notifications dated February 12, 2010. Consignees were informed of the affected product and instructed on proper actions. The firm also asked that the enclosed response forms be returned. For further information, contact your local Beckman Coulter Representative or call 1-800-526-7694.
- Initiated
- 2010-02-04
- Root cause
- Software design
- Terminated
- 2014-09-02
- Distribution
- Distributed Nationwide and Canada.
- Recall number
- Z-1658-2010
- Recall status
- Terminated
- CDRH record ID
- 90221
- FDA classified
- 2010-05-26
- Recall event number
- 55176