LifeSync Wireless ECG System - Intermittent failure when using LS-202 and LS-232 LeadWear in conjunction with LS-41245 and LS-41285 Adaptors. The out of box failure manifests as a V-Lead failure, leading to a possible delay in treatment based upon a delay in ECG monitoring.
Source facts
Action
Firm issued a Device Recall Letter on 12/14/2007. Users were requested to remove the product from inventory and discontinue use of the Large Sized LeadWear with the 12 Lead Adaptors. The current stock of product in inventory will be held under a stop shipment, pending destruction of non-conforming product.
Initiated
2007-12-13
Root cause
Nonconforming Material/Component
Terminated
2008-10-01
Distribution
Class II Recall - Nationwide Distribution.
Recall number
Z-1659-2008
Recall status
Terminated
CDRH record ID
69243
FDA classified
2008-09-16
Recall event number
47336