Open original FDA record ↗Recall record
- FDA identifier
- Z-1659-2011
- Company in FDA record
- Cytyc Prenatal Products/Hologic Inc.
- FDA classification date
- Mar 16, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Product name change requires conversion to a different type of PMA and implies that the product could be used to assess whether a patient's pregnancy is full term, or to predict imminent delivery in women who are at full term gestation.
Source facts
- Action
- The firm, Hologic, sent a "Important-Product Recall/Correction" letter dated December 18, 2009 to its customers. The letter described the product, problem and the actions to be taken. The customers were instructed to 1) replace the directional inserts for the specimen collection kit and health care providers' brochure with provided instructions, 2) over-label specimen collection kit boxes with provided stickers and individual kits, 3) permanently destroy any marketing and educational materials with the "full term" name and 4) complete and return the CUSTOMER RECALL RESPONSE FORM via fax at 408-744-1905 or scan and email to recallfullterm@hologic.com. If you have any questions about this notification, please contact the Hotline (800-442-9892, Option 7# or 508-263-8510).
- Initiated
- 2009-12-15
- Root cause
- PMA
- Terminated
- 2011-03-16
- Distribution
- Worldwide distribution.
- Recall number
- Z-1659-2011
- Recall status
- Terminated
- CDRH record ID
- 87604
- FDA classified
- 2011-03-16
- Recall event number
- 54136