Skip to content

Public FDA record

Z-1659-2025: Radiometer ABL90 Series - Radiometer ABL90 FLEX Model Number 393-090 Radiometer ABL90 Series - Radiomete

Recall record Z-1659-2025 for Radiometer Medical ApS. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1659-2025
Company in FDA record
Radiometer Medical ApS
FDA classification date
Apr 23, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

ABL90 FLEX and ABL90 FLEX PLUS analyzers running with Radiometer Windows 10 versions RWIN10 1.0 and RWIN10 1.1 may unexpectedly freeze (the analyzer ceases to operate) during a sample measurement. The analyzer software includes a feature (watchdog) that detects a freeze and automatically restarts the software to continue operating the analyzer. However, even if a restart occurs, any sample being processed at that time will be lost, potential to lead to delay in test results.

Source facts

Action
On 03/21/2025, the firm sent via email and FedEx an "Urgent: Medical Device Recall" Letter informing customers that, Radiometer has become aware of a potential risk with ABL90 FLEX and ABL90 FLEX PLUS analyzers when operating with Radiometer Windows 10 versions (RWIN10 1.0 and 1.1). Specifically, the analyzer software may unexpectedly "freeze" and restart during a sample measurement, causing the analyzer to cease operation during the startup period. Customers are instructed to: 1. Do not use the affected systems without having an alternative solution readily available. Alternative solutions could include using an additional analyzer or sending the sample to a central laboratory. This also involves ensuring that there is enough patient sample for a second measurement, in case the sample is lost due to the analyzer restarting. Additionally, we ask that you please post a copy of this letter on or near the device(s) until such time that the software has been upgraded. Radiometer will contact customers to schedule the upgrade to analyzers to the available software RWIN10 1.2 or greater. For questions, contact Radiometer at 1-800-736-0600 option 1, Monday - Friday 9 AM to 7 PM (EST).
Initiated
2025-03-21
Root cause
Component design/selection
Distribution
U.S.: AL, AR, AZ, BM CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY. O.U.S.: Canada, Afghanistan, Albania, Algeria, Argentina, Bahrain, Bangladesh, Belarus, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Chile, Colombia, Cuba, Cyprus, Dominican Republic, Ecuador, Egypt, Eritrea, Ethiopia, Greece, Guatemala, Honduras, Iraq, Iran, Israel, Jamaica, Jordan, Kenya, Kuwait, Lebanon, Libya, Lithuania, Malta, Mauritius, Moldova, Nepal, Nicaragua, Nigeria, Oman, Pakistan, Paraguay, Peru, Philippines, Puerto Rico, Qatar, Romania, Saudi Arabia, Serbia, Slovenia, Sri Lanka, Suriname, Syria, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Uganda, Ukraine, United Arab Emirates, Uruguay, Uzbekistan, Venezuela, Vietnam, Yemen, Zimbabwe, Australia, New Zealand, New Caledonia, Tokelau, Austria, Belgium, Netherlands, China, Croatia, Czech Republic, Slovakia, Denmark, Iceland, Finland, Estonia, France, Germany, Hong Kong, Hungary, India, Brunei, Italy, Japan, Kazakhstan, Malaysia, Mexico, Norway, Poland, Russia, Singapore, South Africa, Cameroon, Ivory Coast, Malawi, South Korea, Spain, Portugal, Sweden, Switzerland, Turkey, United Kingdom, Ireland
Recall number
Z-1659-2025
Recall status
Open, Classified
CDRH record ID
213087
FDA classified
2025-04-23
Recall event number
96515

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗