Open original FDA record ↗Recall record
- FDA identifier
- Z-1660-2008
- Company in FDA record
- Lifesync Corporation
- FDA classification date
- Sep 16, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
LifeSync Wireless ECG System - Intermittent failure when using LS-202 and LS-232 LeadWear in conjunction with LS-41245 and LS-41285 Adaptors. The out of box failure manifests as a V-Lead failure, intended for use as a radiofrequency signal transmitter and receiver of electrocardiograph physiological signals which are displayed on the ECG monitors of various manufacturers' systems that have been
Source facts
- Action
- Firm issued a Device Recall Letter on 12/14/2007. Users were requested to remove the product from inventory and discontinue use of the Large Sized LeadWear with the 12 Lead Adaptors. The current stock of product in inventory will be held under a stop shipment, pending destruction of non-conforming product.
- Initiated
- 2007-12-13
- Root cause
- Nonconforming Material/Component
- Terminated
- 2008-10-01
- Distribution
- Class II Recall - Nationwide Distribution.
- Recall number
- Z-1660-2008
- Recall status
- Terminated
- CDRH record ID
- 69244
- FDA classified
- 2008-09-16
- Recall event number
- 47336