Z-1660-2025: Philips Brilliance CT Big Bore, Software Version 4.8.0.10430. Computed Tomography X-Ray System.

Recall record Z-1660-2025 for Philips North America Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1660-2025
Company in FDA record
Philips North America Llc
FDA classification date
Apr 23, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Unintended motion issues related to Interventional Control that may lead to collision of the Gantry/table with the operator or patient.

Source facts

Action
An "URGENT Medical Device Correction" Notice dated 4/1/25 was mailed to consignees. The notice instructs consignees on how to identify if their unit is affected by this recall. Consignees are instructed to continue use of affected devices by following the recommended actions outlined in the notice and Appendix A. Consignees are asked to return the completed URGENT Medical Device Correction Response Form within 30 days of notification of this recall event via email: Philips.Recall@philips.com. The recall notification is to be circulated among users and retained near affected devices until the update is installed. Philips will be contacting consignees to schedule a Field Service Engineer to visit affected sites and install the update (FCO72800825). Consignees with any questions are to call the Customer Care Solutions Center at 1-800-722-9377.
Initiated
2025-04-01
Root cause
Software design
Distribution
Worldwide distribution: US (Nationwide); and OUS (Foreign) to countries of: Algeria, Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Canada, Chile, China, Czech Republic, Estonia, France, Germany, Honduras, Hong Kong, India, Indonesia, Iran, Iraq, Israel, Italy, Lebanon, Libya, Luxembourg, Mauritius, Mexico, Monaco, Morocco, Myanmar, Netherlands, New Zealand, Norway, Palestine, Panama, Philippines, Poland, Portugal, Puerto Rico, R¿union, Russian Fed., Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, T¿rkiye, Turkmenistan, United Kingdom, UAE & Yemen.
Recall number
Z-1660-2025
Recall status
Open, Classified
CDRH record ID
213376
FDA classified
2025-04-23
Recall event number
96667

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗