Open original FDA record ↗Recall record
- FDA identifier
- Z-1661-2010
- Company in FDA record
- Ge Healthcare, Llc
- FDA classification date
- May 26, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
A potential software failure related to the Worklist functionality has been identified on a subset of Innova 2100IQ, 3100IQ and 4100IQ systems. When using the Worklist function to import patient data to Innova, there are two scenarios that will cause the next exam to fail and the inability to recall acquired sequences.
Source facts
- Action
- Consignees were sent on 3/24/10, a GE Healthcare " Urgent Medical Device Correction" letter dated March 23, 2010. The letter included the Safety Issue, Affected Product Details, Product Correction and Contact Information.
- Initiated
- 2010-03-24
- Root cause
- Software design
- Terminated
- 2013-05-21
- Distribution
- Worldwide distribution: USA, Puerto Rico, VENEZUELA, UNITED KINGDOM, TURKEY, THAILAND, TAIWAN, PROVINCE OF CHINA, SWEDEN, SPAIN, RUSSIAN FEDERATION, ROMANIA, POLAND, PHILIPPINES, NORWAY, NETHERLANDS, MOROCCO, REPUBLIC OF KOREA, KAZAKSTAN, JAPAN, ITALY, ISLAMIC REPUBLIC OF IRAN, INDONESIA, INDIA, GERMANY, FRANCE, FINLAND, EGYPT, CHINA, CHILE, CANADA, BRAZIL, BELGIUM, BELARUS, AUSTRALIA, and ARGENTINA.
- Recall number
- Z-1661-2010
- Recall status
- Terminated
- CDRH record ID
- 90872
- FDA classified
- 2010-05-26
- Recall event number
- 55445