Open original FDA record ↗Recall record
- FDA identifier
- Z-1663-2023
- Company in FDA record
- TELEFLEX LLC
- FDA classification date
- May 30, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Teleflex LLC is initiating this voluntary field corrective action because the clip cartridges for the affected product is missing a key feature that restricts the application of the clip to the applier. With this step feature missing on the cartridge, the clip is mounted in a more variable position, allowing the clips to load farther back in the clip applier jaw grooves. This may potentially result in inappropriate or inadequate clip application to vascular structures.
Source facts
- Action
- Teleflex issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 04/28/2023 by FedEx 2nd day mail. The letter explained the issue, potential risk, and requested the following actions: Medical facilities: Cease use and distribution, and isolate the affected product. Circulate the notice to all affected departments. Distributors: Cease use and distribution, and isolate the affected product. Provide the notice to your customers. Teleflex is seeking return of the affected units.
- Initiated
- 2023-04-28
- Root cause
- Under Investigation by firm
- Distribution
- worldwide
- Recall number
- Z-1663-2023
- Recall status
- Open, Classified
- CDRH record ID
- 199907
- FDA classified
- 2023-05-30
- Recall event number
- 92212