Z-1663-2025: Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount

Recall record Z-1663-2025 for IHB OPERATIONS B.V.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1663-2025
Company in FDA record
IHB OPERATIONS B.V.
FDA classification date
Apr 23, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column.

Source facts

Action
Skytron issued a Field Correction notice to its service personnel on 03/27/2025 via email. The notice explained the issue, risk to health, and requested the following: Customer Acknowledgement 1. Contact customers to ensure they have received the customer letter. 2. Provide a copy of the customer letter if needed. 3. Document who you spoke with and the date of contact on the Customer Acknowledgement Spreadsheet. Customer Acknowledgement Spreadsheet will be provided by Quality. 4. Return the completed Customer Acknowledgement Spreadsheet to quality@skytron.com. The devices will be corrected on site. Skytron notified the medical facilities (users of the device) by issuing an URGENT: MEDICAL DEVICE RECALL notice on 03/28/2025 via US Mail. This letter explained the issue, risk to health and requested the following: Actions to be taken by Customer/User: Prior to the field correction, you should take the following precautionary measures to mitigate any potential risks: Inspect your 4FXS-60 frames to ensure that all 4 mounting screws are secure. If any loose connections are identified, immediately remove the 4FXS-60 from use and notify your local authorized representative. For questions or concerns, please contact your local authorized representative or Skytron at 1-800-759-8766.
Initiated
2025-03-27
Root cause
Process change control
Distribution
US Nationwide distribution including in the states of AL, AZ, CA, FL, IA, ID, IL, LA, MA, MD, MI, MN, NC, NH, NJ, NM, NY, PA, TN, TX, UT, VA, WI, WV.
Recall number
Z-1663-2025
Recall status
Open, Classified
CDRH record ID
213205
FDA classified
2025-04-23
Recall event number
96581

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 21, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗