- FDA identifier
- Z-1664-2023
- Company in FDA record
- Ortho-Clinical Diagnostics, Inc.
- FDA classification date
- May 30, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Unavailable to calibrate VITROS Chemistry Products VANC Reagent, Generation (GEN) 50, results in delay of testing
Source facts
- Action
- QuidelOrtho (formerly Ortho Clinical Diagnostics) issued Urgent Product Correction Notification Letter on 4/14/23 to Distributors and End/Users. Letter states reason for recall, health risk and action to take: Prior to calibrating VITROS VANC Reagent GEN 50 using VITROS Calibrator Kit 11, Lot 1172, install ADD DRV 6224 or above on your VITROS System(s). " Complete the enclosed Confirmation of Receipt form no later than April 24, 2023. " Forward this notification if VITROS Calibrator Kit 11, Lot 1172 was distributed outside of your facility. " Save this notification with your user documentation or post this notification by each VITROS 5,1 FS/4600/5600/XT 7600 System until ADD DRV 6224 or above has installed. " If your laboratory has experienced the issue with this product and you have not already done so, please report the occurrence to your local Ortho Care" Technical Solutions Center. Contact Information If you have further questions, please contact the Ortho Care Technical Solutions Center at 1-800-421-3311.
- Initiated
- 2023-04-14
- Root cause
- Under Investigation by firm
- Distribution
- Nationwide Foreign: Brazil, Chile, Colombia
- Recall number
- Z-1664-2023
- Recall status
- Open, Classified
- CDRH record ID
- 199984
- FDA classified
- 2023-05-30
- Recall event number
- 92245