- Recall classification
- Class II
Reason stated in the source
When DxI 9000 Access Immunoassay Analyzer is connected to a host system (laboratory information system (LIS) or middleware) and has accumulated canceled QC test results that were not sent by LIS, the analyzer may lose communication with the host system. The lost connection interrupts sample processing, which delays reporting patient test results and may subsequently delay patient treatment.
Source facts
- Action
- Beckman Coulter issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 03/27/2025 via USPS first class mail. The notice explained the issue with the device, impact to the patient, and provided a workaround until a software upgrade is available. If the consignee further distributed the affected device, they were directed to provide a copy of the notice to the recipient. For questions, contact Customer Support Center, from the website: http://www.beckmancoulter.com
- Initiated
- 2025-03-27
- Root cause
- Software design
- Distribution
- Worldwide distribution.
- Recall number
- Z-1664-2025
- Recall status
- Open, Classified
- CDRH record ID
- 213154
- FDA classified
- 2025-04-24
- Recall event number
- 96554