Open original FDA record ↗Recall record
- FDA identifier
- Z-1666-2019
- Company in FDA record
- Maquet Cardiovascular, LLC
- FDA classification date
- May 31, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured
Source facts
- Action
- Maquet Getinge issued recall notification letter on April 18, 2019 via FedEx 2-day delivery identifies issue, health risk and action to take: quarantine and return to Maquet/Getinge recalled product. Complete response form and return to acknowledge this recall. Questions contact Getinge Customer Service at 1 888 627 8383 (press option 2, then option 2) between the hours of 8 AM and 6 PM Eastern Standard Time. On 3/26/20 firm expanded recall to include an additional lot for one consignee,product: "BEQ TOP 25003 3/8 3/8 up to 7 lpm", item number 701063255R02, lot 3000087051.
- Initiated
- 2019-04-18
- Root cause
- Device Design
- Terminated
- 2025-07-28
- Distribution
- Nationwide
- Recall number
- Z-1666-2019
- Recall status
- Terminated
- CDRH record ID
- 172839
- FDA classified
- 2019-05-31
- Recall event number
- 82703