Open original FDA record ↗Recall record
- FDA identifier
- Z-1666-2025
- Company in FDA record
- Remote Diagnostic Technologies Ltd.
- FDA classification date
- Apr 24, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
software issue that causes an error screen on a portable vital signs monitor when attaching video Laryngoscope.
Source facts
- Action
- On April 16, 2025, Remote Diagnostic Technologies Ltd (subsidiary of Philips) issued a Urgent Medical Device Correction to affected consignees via mail. Philips ask consignees to take the following: 1. DO NOT connect the Video Laryngoscope USB probe to the Tempus Pro during device boot-up. Once Patient Details dialog appears, enter patient data, then plug in the Video Laryngoscope USB probe. 2. If the Tempus Pro Monitor displays the error dialog screen (where the user is asked to restart the device), press the Shutdown Button and restart the Tempus Pro Monitor. 3. Keep a copy of this letter with or near your Tempus Pro Monitor. 4. Complete and return the Urgent Field Safety Notification Response Form included with this letter within 30 days of receipt of this notice. 5. This notice must be shared with all relevant personnel within your organization and with any organization where the potentially affected devices have been transferred.
- Initiated
- 2025-04-16
- Root cause
- Device Design
- Distribution
- Worldwide - US Nationwide distribution including in the states of AL, AZ, CA, CO, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, WI, WV, WY and the countries of AE, AF, AT, AU, BE, BO, CA, CH, CZ, DE, DK, ES, FI, FR, GB, HK, IE, IL, IS, IT, KY, MY, NL, NO, OM, PL, PT, RO, SA, SE, SG, SI, SL, TH, ZA.
- Recall number
- Z-1666-2025
- Recall status
- Open, Classified
- CDRH record ID
- 213391
- FDA classified
- 2025-04-24
- Recall event number
- 96678