Philips Medical Systems North America Co. Phillips
FDA classification date
Sep 24, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification
Reason stated in the source
The x-ray system fail to meet the requirements of 21 CFR 1020.32(d)(2)(ii) when a specific tube failure occurs involving fluoroscopy of large patients and a partial failure of the grid switch
Source facts
Action
A Customer Notification letter describing the potential grid switch failure and actions planned by Philips to correct the problem will be sent to all users of the affected systems during the week of July 6, 2009.
Initiated
2009-07-06
Root cause
Radiation Control for Health and Safety Act
Terminated
2011-12-16
Distribution
Worldwide Distribution
Recall number
Z-1667-2009
Recall status
Terminated
CDRH record ID
84493
FDA classified
2009-09-24
Recall event number
53034
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.