Z-1667-2023: NOxBOXi Nitric Oxide Delivery System, NOXBOX-I, REQNOXBOX-I, Nitric Oxide Delivery Device

Recall record Z-1667-2023 for NOXBOX LTD. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1667-2023
Company in FDA record
NOXBOX LTD
FDA classification date
Jun 9, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Device may experience malfunctions due to misalignment of the check valve in the manifold. If the device fails, there is a risk of nitric oxide leak, oxygen gas leak, or earlier cylinder gas changeover than expected or not at all. In a worst case scenario, there may be an interruption in therapy which could lead to oxygen desaturation of the patient and associated complications.

Source facts

Action
On May 3, 2023, the firm notified customers via URGENT MEDICAL DEVICE RECALL letters. Devices do not need to be removed from service unless they do not pass the start-up high-pressure leak test or it alarms during cylinder changeover and troubleshooting does not resolve the alarms. Customers should prioritize use of a device that is not affected, if available, when initiating therapy on a new patient. If there are no unaffected devices available, or if an affected device is already in use on a patient, follow troubleshooting measures as provided in the letter. Once new manifolds are available, NBL will correct or replace affected devices on a rolling basis. If you have any questions or require further assistance, please contact your sales representative or customer service at 833-NOX-VENT.
Initiated
2023-05-03
Root cause
Component change control
Terminated
2024-10-31
Distribution
Worldwide distribution - US Nationwide and the countries of Bahrain, Bolivia, Brazil, Ecuador, Finland, France, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Kuwait, Malaysia, Mexico, Netherlands, New Zealand, Panama, Saudi Arabia, Singapore, South Africa, Spain, Taiwan, Turkey, UAE, United Kingdom.
Recall number
Z-1667-2023
Recall status
Terminated
CDRH record ID
199818
FDA classified
2023-06-09
Recall event number
92171

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗