Affected pumps may trigger an erroneous (false) Upstream Occlusion Alarm under certain conditions, which will interrupt an active infusion. Interruption or delay of therapy can lead to serious patient injury or death.
Source facts
Action
On April 10, 2025, the firm notified affected consignees via letters titled "Urgent Medical Device Correction."
Customers were informed of conditions under which the false Upstream Occlusion Alarm may occur. The USO alarm can be cleared during these conditions by removing the administration from the pump, re-attaching the administration set, and restarting delivery. Customers should verify there is no upstream occlusion present.
The firm will be making an update to the Infusion Pump Operator's Manual to include these conditions.
Initiated
2025-04-10
Root cause
Labeling design
Distribution
Worldwide - US nationwide distribution.
Recall number
Z-1667-2025
Recall status
Open, Classified
CDRH record ID
213267
FDA classified
2025-05-07
Recall event number
96620