Open original FDA record ↗Recall record
- FDA identifier
- Z-1668-2014
- Company in FDA record
- Siemens Medical Solutions USA, Inc.
- FDA classification date
- May 28, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential malfunction due to a software issue for ACUSON SC2000 ultrasound systems using software version 3.5 (VA35) when using the stress echo feature which could cause the need for the study to be repeated.
Source facts
- Action
- Siemens sent a Consumer Safety Advisory Notification letter on April 23, 2014, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discontinue use of the stress echo feature with software version .5. For studies acquired with systolic-only captures: review all patient studies on exams performed after software revision 3.5 was installed on your system. If you are unsure when this software was installed on your system, please contact your local Customer Service representative. Foreign distributors will implement distribution of the field correction to their customers. For questions regarding this recall call 650-694-5398.
- Initiated
- 2014-04-23
- Root cause
- Software design
- Terminated
- 2015-11-12
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to Australia, Belarus, Dominican Rep., France, Germany, India, Indonesia, Israel, Japan, Mexico, New Zealand, Norway, Philippines, Poland, Republic of Korea, Romania, Russian Fed., Saudi Arabia, Singapore, Sweden, Switzerland, Turkey, U.A.E., Ukraine, and United Kingdom.
- Recall number
- Z-1668-2014
- Recall status
- Terminated
- CDRH record ID
- 127127
- FDA classified
- 2014-05-28
- Recall event number
- 68197