Skip to content

Public FDA record

Z-1668-2023: Intelli-C, Right side suspended Tabletop, REF: 03400000

Recall record Z-1668-2023 for NRT X-RAY A/S. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1668-2023
Company in FDA record
NRT X-RAY A/S
FDA classification date
May 31, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential that the drive shaft in the detector up/down (IDE) movement of the X-Ray machine may break which causes the detector to move until it is stopped by the movement end-stop.

Source facts

Action
Between the dates of April 27, 2023 - May 1, 2023, NRT X-Ray A/S issued a "urgent: Medical Device Correction" via E-Mail to affected consignees. Per the customer communication, NRT ask consignees to take the following actions: Step 1 Short term action The first step of the recall is a corrective action which requires the existing fixation bracket components to be removed. Service Providers of the Intelli-C systems should follow the instructions in page 4 through 7 of the communication and report back to NRT using the Customer Response Card on page 8. The process is coordinated through US dealer Alpha Imaging in Willoughby, Ohio. Step 2 Long term action The second step is a preventive action to replace the IDE drive shaft and relating components. A separate set of instructions for this action will be provided in May together with the required parts for the replacement. These instructions will include a separate acknowledgement form for this action. If you have questions regarding the contents of this letter, please contact your local Service provider or contact NRT X-RAY A/S directly using the following contact information: Email: aqr@nrtxray.com Phone: +45 2055 3309 Monday through Friday, (03:00)-11:00 Eastern Standard Time (9:00 AM to 5:00 PM, Central European Time)
Initiated
2023-04-27
Root cause
Device Design
Distribution
US Nationwide distribution in the states of OH, NC, KY, SC, IA, MA, WA, LA, NV, NY.
Recall number
Z-1668-2023
Recall status
Open, Classified
CDRH record ID
199949
FDA classified
2023-05-31
Recall event number
92230

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗