Skip to content

Public FDA record

Z-1670-2013: Sovereign Compact Phacoemulsification System, v 5.3, Part Number: CMP680300. The system is used for catar

Recall record Z-1670-2013 for Abbott Medical Optics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1670-2013
Company in FDA record
Abbott Medical Optics, Inc.
FDA classification date
Jul 9, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Abbott Medical Optics Inc. (AMO) initiated this Urgent Notice of Medical Device Correction and Advisory Notice to all customers who use the AMO Sovereign Compact Phacoemulsification System, Version 5.3, because it has the potential to lose the vacuum tone mode during surgery.

Source facts

Action
AMO issued an "Urgent Notice of Medical Device Correction and Advisory Notice" on May 23, 2013 via email, to all customers with a Sovereign Compact Phacoemulsification System v 5.3 to inform them of the issue and provide them with actions to take in the event of sound loss during use of the Sovereign Compact Phacoemulsification System, v 5.3.
Initiated
2013-05-23
Root cause
Software design
Terminated
2014-10-06
Distribution
Worldwide Distribution, including Nationwide (US) and countries in Latin America, EMEA, and APAC.
Recall number
Z-1670-2013
Recall status
Terminated
CDRH record ID
118560
FDA classified
2013-07-09
Recall event number
65292

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗