Abbott Medical Optics Inc. (AMO) initiated this Urgent Notice of Medical Device Correction and Advisory Notice to all customers who use the AMO Sovereign Compact Phacoemulsification System, Version 5.3, because it has the potential to lose the vacuum tone mode during surgery.
Source facts
Action
AMO issued an "Urgent Notice of Medical Device Correction and Advisory Notice" on May 23, 2013 via email, to all customers with a Sovereign Compact Phacoemulsification System v 5.3 to inform them of the issue and provide them with actions to take in the event of sound loss during use of the Sovereign Compact Phacoemulsification System, v 5.3.
Initiated
2013-05-23
Root cause
Software design
Terminated
2014-10-06
Distribution
Worldwide Distribution, including Nationwide (US) and countries in Latin America, EMEA, and APAC.
Recall number
Z-1670-2013
Recall status
Terminated
CDRH record ID
118560
FDA classified
2013-07-09
Recall event number
65292