Study and all acquired images deleted when using systems with software version VE10E. Acquisitions have to be repeated. Connection with portable detectors (MAX wi-D, MAX mini) is sporadic in image recovery. Pressing undo button on the RAD subtask card or previous series button on the Image subtask card during image readout. Sporadically, during an automatic or a manual RIS update.
Source facts
Action
A safety advisory notice, dated April 13, 2015, was sent to end users with affected devices to alert them about the issue and associated risks,and to provide instructions for avoiding potential risk. A new software version will be released to permanently resolve the issue.
Initiated
2015-04-13
Root cause
Software design
Terminated
2017-03-20
Distribution
Nationwide Distribution.
Recall number
Z-1670-2015
Recall status
Terminated
CDRH record ID
137253
FDA classified
2015-05-23
Recall event number
71115