Z-1672-2021: GS777 Wall Transformer Model: 77710 Intended Use: The GS777 Wall Transformer is intended for use

Recall record Z-1672-2021 for Welch Allyn Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1672-2021
Company in FDA record
Welch Allyn Inc
FDA record creation date
May 21, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential Risk of fluid ingress to the power supplies, resulting in risk of an internal electrical short and possible electrical shock

Source facts

Action
Welch Allyn, Inc. (a Hill-Rom Company) issued Customer Medical Device Correction letter ( FSCA-identifier: MOD1329) mailed (with signature service) on April 22, 2021(with signature service) via third Party. The letter states reason for recall, health risk and risk mitigations until the updated Instructions for Use (IFU) becomes available. To minimize risk of potential electrical shock be sure to follow these cleaning recommendations: When Cleaning is performed: 1. Unplug the power supply from the outlet. 2. Ensure the cleaning cloth is not soaking (not dripping or over saturated) with cleaning fluid. 3. Apply minimal pressure on wipe when cleaning; avoid wiping gap areas identified on adapter plug as these areas are most susceptible to liquid ingress. 4. Ensure the power supply and power cord are dry before plugging into electrical outlet. 5. Hold the sides of the power supply when plugging into electrical outlet as pictured below. Avoid contact with bottom of the power supply where cleaning fluid is mostly likely to have accumulated The customer will be instructed to sign acknowledgement form to confirm receipt and return to a dedicated email address. Distributors will be asked to forward to their end users. They can request an electronic copy of the FSN and response form to share with end users until updated IFU becomes available. Distributors will be provided instruction on communication with their end users once response form is received. Consignees of new sales will receive the Field Safety Notice until the IFU is updated. This will be done by pulling updated consignee lists monthly and providing any new consignees with the customer letter. .
Initiated
2021-04-22
Root cause
Device Design
Terminated
2024-05-31
Distribution
Worldwide - US Nationwide distribution Foreign: Afghanistan Albania Algeria American Samoa Argentina Australia Austria Azerbaijan Bahamas (the) Bahrain Bangladesh Barbados Belgium Bermuda Bhutan Bolivia (Plurinational State of) Botswana Brazil Brunei Darussalam Bulgaria Cameroon Canada Chad Chile China Colombia Congo (the Democratic Republic of the) Congo (the) Costa Rica Croatia Cyprus Czechia Denmark Dominican Republic (the) Ecuador Egypt El Salvador Estonia Eswatini Finland France French Polynesia Georgia Germany Ghana Greece Guam Guatemala Honduras Hong Kong Hungary Iceland India Indonesia Iran (Islamic Republic of) Iraq Ireland Israel Italy Jamaica Japan Jordan Kenya Korea (the Republic of) Kuwait Lao People's Democratic Republic (the) Latvia Lebanon Lesotho Libya Luxembourg Malaysia Maldives Malta Mexico Mozambique Myanmar Namibia Nepal Netherlands (the) New Zealand Nicaragua Nigeria Northern Mariana Islands (the) Norway Oman Pakistan Panama Paraguay Peru Philippines (the) Poland Portugal Puerto Rico Qatar R¿union Romania Russian Federation (the) Saudi Arabia Senegal Serbia Singapore Slovenia South Africa Spain Sri Lanka Sweden Switzerland Taiwan (Province of China) Thailand Trinidad and Tobago Tunisia Turkey Uganda Ukraine United Arab Emirates (the) United Kingdom Venezuela (Bolivarian Republic of) Viet Nam
Recall number
Z-1672-2021
Recall status
Terminated
CDRH record ID
186902
FDA record created
2021-05-21
Recall event number
87749

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗