Open original FDA record ↗Recall record
- FDA identifier
- Z-1673-2009
- Company in FDA record
- Volk Optical Inc
- FDA classification date
- Jul 14, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Lens may fall onto eye surface during procedure. Ophthalmologic device had a loose set screw on the Ball-Seal bushing assembly. As a result, the set screw becomes disengaged from the set screw groove in the bushing assembly and the Lens Positioning Unit slides downward.
Source facts
- Action
- Volk Optical sent a 'MEDICAL DEVICE RECALL' letter to their sole customer on 1/23/2009. The letter requests the return of the remaining two (2) units in the customer's possession (this firm experienced the device failure and reported the consumer complaint which prompted the recall; and had already returned the affected unit to the Volk for refund or replacement). The letter provides Volk's Fedex and UPS numbers for the customer to use to return the recalled units and also provides the E-mail address for the firm's contact person concerning this recall should additional assistance be required.
- Initiated
- 2009-01-23
- Root cause
- Process design
- Terminated
- 2012-04-12
- Distribution
- USA distribution to a single customer located in MO.
- Recall number
- Z-1673-2009
- Recall status
- Terminated
- CDRH record ID
- 83084
- FDA classified
- 2009-07-14
- Recall event number
- 52461