The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly state the level at which biotin does not interfere.
Source facts
Action
Siemens Healthcare Diagnostics issued a recall in March 2017 for products with incorrect units for biotin in the Instructions for Use. After further investigation, recalling firm determined that more Siemens products also had missing or incorrect information in the IFUs for which two recalls were initiated, and Amended Corrections and Removals Reports were submitted to the FDA
Initiated
2018-01-02
Root cause
Error in labeling
Terminated
2023-02-10
Distribution
Nationwide and Canada, Mexico, and Thailand
Recall number
Z-1673-2018
Recall status
Terminated
CDRH record ID
162503
FDA classified
2018-05-04
Recall event number
79505