Z-1673-2020: LabPro Data Management System, UDI: 15099590676872 - Product Usage: The LabPro Data Management System is

Recall record Z-1673-2020 for Beckman Coulter Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1673-2020
Company in FDA record
Beckman Coulter Inc.
FDA record creation date
Apr 6, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A security-only update was released via remote diagnostics on January 14, 2020 to customers with Windows 7 operation system. Shortly after the releases some customers reported that Data Management System computers would not restart. If the computer is unable to restart, then the system cannot download data generated from MicroScan instruments or manually entered microbiology test results, resulting in a potential of delayed results.

Source facts

Action
On January 17, 2020 Beckman Coulter sent an "Urgent Medical Device Recall" email to all affected customers. In addition to informing customers about the recall they asked customers to take the following action: If you have received a prompt to restart the LabPro computer on or after January 14, 2020 and have not yet restarted the computer, then do not restart the computer and you will be contacted by your Beckman Coulter Representative for further instructions. If you are unsure if you have received the restart prompt, then monitor the LabPro computer for at least an hour to see if the restart prompt appears. The prompt will redisplay hourly, if prompted, do not restart the computer. If you have received a prompt to restart the LabPro computer on or after January 14, 2020, and successfully restarted the computer, then no further action is required. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected products listed above to another laboratory, please provide them a copy of this letter. In order to assure you have received this important communication, please respond within 10 days in one of the following ways: - Electronically, if you received this communication via email. - Manually, complete and return the enclosed Response Form. If you have any questions regarding this notice, please contact our Customer technical Support Center: - Via our website, http://www.beckmancoulter.com/customersupport/support - Via phone, call 1-800-677-7226 in the United States - If outside the United STates, please contact your local Beckman Coulter Representative
Initiated
2020-01-17
Root cause
Software Manufacturing/Software Deployment
Terminated
2021-04-27
Distribution
Worldwide distribution - US Nationwide distribution in the states of Wisconsin, New York, Delaware, Indiana, Ohio, Tennessee, Florida, Georgia, Illinois, South Carolina, California, New Jersey, Alabama, Massachusetts, Michigan, Texas, West Virginia, Wyoming, Missouri, New Mexico, North Carolina, Minnesota, New Hampshire, Oklahoma, Washington, Pennsylvania, Idaho, Mississippi, Maryland, Virginia, Arizona, Utah, Maine, Michigan, Colorado, Iowa, Louisiana, Kentucky, South Dakota, Nevada, North Dakota, Kentucky and countries of Brazil, Germany.
Recall number
Z-1673-2020
Recall status
Terminated
CDRH record ID
179968
FDA record created
2020-04-06
Recall event number
85048

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗