Skip to content

Public FDA record

Z-1673-2022: Pentacam AXL REF 70100, Pentacam HR REF 70900. Used to image the anterior segment of the eye.

Recall record Z-1673-2022 for Oculus Optikgeraete GMBH. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1673-2022
Company in FDA record
Oculus Optikgeraete GMBH
FDA record creation date
Sep 2, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a software issue, IOL calculator printout often does not accurately reflect the alignment axis and incision position when planning toric IOLs

Source facts

Action
On 07/08/2022, OCULUS GmbH issued an "Urgent safety information Corrective Action " Letter to its subsidiary OCULUS Inc via UPS express which identifies a software error in the software version 1.26r26 and 1.26r27 in the IOL calculator of the Pentacam software. Corrective measures to be taken: A software update to version 1.26r28 must be performed on all computers used and affected in connection with the Pentacam and additional workstations. Users should refrain from using the printout of the IOL calculator for affected installations until after the software update has been carried out. For questions or support - contact OCULUS Inc Service Team at 425-670-9977 or email service@oculususa.com or fsca@oculus.de
Initiated
2022-07-08
Root cause
Software design
Distribution
U.S.: AL, AR, AZ, CAQ, CO, CT, FL, GA, HI, IA, ID, IL, KS, KY, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NY, OH, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, WA, WI, and WY O.U.S.: N/A
Recall number
Z-1673-2022
Recall status
Open, Classified
CDRH record ID
195360
FDA record created
2022-09-02
Recall event number
90661

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗