Z-1674-2020: Beckman Coulter MicroScan PROMPT Inoculation System-D, Catalog Number B1026-10D, UDI: 15099590659561 - Pr

Recall record Z-1674-2020 for Beckman Coulter Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1674-2020
Company in FDA record
Beckman Coulter Inc.
FDA classification date
Apr 6, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to customer reports it has been determined that the inoculation system contains the incorrect prompt wands. Use of the incorrect wands could result in false -susceptible results, false-resistant results, possible misidentifications, or possible out-of-range Quality Control results.

Source facts

Action
On 12/13/2019, the firm notified its customers via "Urgent Medical Device Recall" letter informing customers of the recall and its potential to cause false-susceptible results, false-resistant results, possible misidentifications, or possible out-of-range Quality Control results. The notification instructs customers to take the following actions: 1. Discontinue use and discard any remaining inventory of the incorrect wands with lot number 13.LAG.1 that may be found with the Prompt Inoculation System-D lot number 20201103T. 2. The notification letter also instructs customers to share the contents of the letter with lab/medical directors to determine the need to review previous patient and quality control test results. 3. In order to assure you have received the important communication, please respond within 10 days in one of the following ways: - Electronically, if you received this communication via email - Manually, complete and return the enclosed Response Form 4. If you have any questions regarding this notice, please contact the firm's Customer Support Center: - Via our website, http://www.beckmancoulter.com/customersupport/support - Via phone, call 1-800-677-7226 in the US 5. For customers in the US, if you need replacement products: - Complete the attached "Replacement Order Form" and email the form to askbeckman@beckman.com or fax to (866) 294-7850 or Call Client Services at (800) 526-3821
Initiated
2019-12-13
Root cause
Nonconforming Material/Component
Terminated
2021-04-30
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL, AZ, AR, CA, CO, CT, DE, FL, GA, IL, IN, IA, KS, KY, LA, MD, MI, MS, NE, NY, NC, ND, OH, OK, PA, SD, TN, TX, UT, WV, WI, WY and countries of Canada, China, Japan, Paraguay.
Recall number
Z-1674-2020
Recall status
Terminated
CDRH record ID
180153
FDA classified
2020-04-06
Recall event number
85137

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗