Z-1674-2021: Centaur Syphilis Assay-SYP Kit, 200 Test - Product Usage: For in vitro diagnostic use in the qualitative

Recall record Z-1674-2021 for Siemens Healthcare Diagnostics, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1674-2021
Company in FDA record
Siemens Healthcare Diagnostics, Inc
FDA record creation date
May 24, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for the ADVIA Centaur Syphilis assay to carryover into other commercial assays. This carryover effect only occurs when the impacted assays are used immediately following testing with the Syphilis assay on the ADVIA Centaur XP/XPT or ADVIA Centaur CP systems

Source facts

Action
Siemens issued Urgent Medical Device Correction (UMDC) to US affected customers via FedEx beginning on 04/21/2021. An Urgent Field Safety Notice (UFSN) was issued to all OUS Siemens sites with affected customers electronically for distribution and regional reporting to their local competent authorities beginning on 04/21/2021. Letter states reason for recall, health risk and Actions to be Taken by the Customer: Please review this letter with your Medical Director. If you have multiple ADVIA Centaur XP/XPT or ADVIA Centaur CP systems: Siemens recommends testing the ADVIA Centaur Syphilis assay on a separate system from the impacted ADVIA Centaur assays (ADVIA Centaur PRGE, T4 on the ADVIA Centaur XP/XPT and ADVIA Centaur AFP, COR, PHTN, VALP on the ADVIA Centaur CP) to avoid the potential carryover effect. If you have a single ADVIA Centaur XP/XPT or ADVIA Centaur CP system: It is important to prevent the impacted assays from being processed after the Syphilis assay without a Daily Cleaning Procedure (DCP) run in between. For example, this can be achieved by processing all Syphilis testing, running the DCP, and then proceeding with all other testing including the impacted assays. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. If you have received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. Please retain this letter with your laboratory records and forward this letter to those who may have received this product.
Initiated
2021-04-21
Root cause
Device Design
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Austria, Belgium, Burkina Faso, Croatia, Czech Republic, Egypt, Finland, France, Fren.Polynesia, Georgia, Germany, Germany, Greece, Guadeloupe, Italy, Kazakhstan, Latvia, Netherlands, Norway, Poland, Portugal, Qatar, Russian Fed., Serbia, Slovakia, Spain, Sweden, Switzerland, Turkey, U.A.E., Ukraine, United Kingdom.
Recall number
Z-1674-2021
Recall status
Open, Classified
CDRH record ID
186970
FDA record created
2021-05-24
Recall event number
87770

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗