Support mechanism may not engage properly due to improper stowage after opening which may result in internal component damage or potential injury.
Source facts
Action
The firm has issued a modification of the telescope support arm parking bracket via Update Instruction CT028/09/S. A firm service representative will install the modified support arm parking brackets on the affected systems.
Initiated
2009-05-29
Root cause
Equipment maintenance
Terminated
2009-09-01
Distribution
Nationwide Distribution -- including states of AR, AZ, CA, CO, CT, FL, IA, IL, IN, KS, KY, MA, MD, MN, MO, MT, NC, ND, NE, NH, NY, OH, PA, SC, SD, TN, VA, WA, WI, and WY.
Recall number
Z-1675-2009
Recall status
Terminated
CDRH record ID
82993
FDA classified
2009-07-30
Recall event number
52435
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.