Top table lateral movement on their DUA-450F Kalare Table cannot be performed, or the table top lateral movement is performed at an extremely low speed. As a result, positioning of the patient cannot be performed by tabletop movement.
Source facts
Action
Toshiba America Medical Systems (TAMS) issued Urgent Medical Device Correction notification dated July 8, 2010, to consignees via USPS return receipt mail. The letter contains a return reply form that is to be faxed to TAMS as proof of notification.
Initiated
2010-07-15
Root cause
Software design
Terminated
2012-04-20
Distribution
US Nationwide
Recall number
Z-1675-2011
Recall status
Terminated
CDRH record ID
93589
FDA classified
2011-03-16
Recall event number
56415