Z-1675-2013: AngioDynamics RITA Main Cable, 9 Pin to 14 Pin, REF 700-101892. RITA Main Cables are indicated for use wi

Recall record Z-1675-2013 for Angiodynamics, INC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1675-2013
Company in FDA record
Angiodynamics, INC
FDA classification date
Jul 10, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The RITA Main Cable product contains the incorrect Instructions for Use.

Source facts

Action
Consignees were notified via certified mail on 06/11/2013 with an "Urgent Medical Device Field Safety Correction" notification. This included the correct Instructions for Use and the recommended actions.
Initiated
2013-06-11
Root cause
Packaging process control
Terminated
2015-06-12
Distribution
Worldwide Distribution, including the US, Austria, China, Italy, Russia and the UK.
Recall number
Z-1675-2013
Recall status
Terminated
CDRH record ID
118894
FDA classified
2013-07-10
Recall event number
65439

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗