Software anomaly in the Panorama System software version 8.9 that manifests when the Panorama is in use with Mindray Passport V Monitor. Arrhythmia alarms which were previously displayed on both the Passport V and Panorama will not be displayed on the Panorama following a restart of either system (due to communication loss, restart, etc.).
Source facts
Action
Mindray sent a Corrective Action letter dated March 12, 2015 to their affected customer via FedEx. The letter identified the affected product, problem and actions to be taken. For questions call 201-995-8407.
Initiated
2015-03-12
Root cause
Software design
Terminated
2017-03-28
Distribution
Worldwide Distribution - US Nationwide in the state of TX.
Recall number
Z-1675-2015
Recall status
Terminated
CDRH record ID
135725
FDA classified
2015-05-26
Recall event number
70959