Z-1675-2023: Elekta Unity Image-Guided Radiation Therapy System containing an MR gradient coil.

Recall record Z-1675-2023 for Elekta, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1675-2023
Company in FDA record
Elekta, Inc.
FDA classification date
Jun 2, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a low possibility that an electrical connector in the MR gradient coil will overheat on Elekta Unity systems.

Source facts

Action
The recalling firm issued the Field Safety Notice dated March 2023 via email on 4/6/2023 describing the problem, the clinical impact, and the recommended user action, which is that no actions are required to be taken by the user. The Unity system is safe for clinical use without restriction. The consignee is informed the letter is to notify them there will be action taken on the Elekta Unity system to further reduce the potential risk of overheating within the MR Gradient Coil which could lead to system downtime. The letter is to be posted in a place accessible to all users and appropriate personnel are to be advised on the content of the letter. The letter contains Elekta Corrective Actions, which says the update will be provided via Elekta's Field Change Order process once available and that their local Elekta service representative will contact the consignee to arrange for this activity to be completed. An acknowledgement Form was enclosed to be returned to Elekta immediately upon receipt but no later than within 30 days. The device serial number and location or site of the device is to be reported and the consignee is to sign it to acknowledge they have read and understand the notice and accept the implementation of any given recommendation. UPDATED 12-JUN-23 An updated Important Field Safety Notification was released to customers by email in which supersedes the previous Notice under ref: 200-01-801-012 released on 06-APR-23. The new FSN includes additional information under Recommended User Action.
Initiated
2023-04-06
Root cause
Process design
Distribution
US Nationwide distribution in the states of FL, IA, NJ, NY, OK, PA, TN, WI, MI, LA, KS, NM and TX.
Recall number
Z-1675-2023
Recall status
Open, Classified
CDRH record ID
199492
FDA classified
2023-06-02
Recall event number
92053

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗