Z-1675-2025: Straight Inflow/Outflow Patient Line Kit. Model Number: 902-00037. Sterile, single-use, disposable set

Recall record Z-1675-2025 for Belmont Instrument LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1675-2025
Company in FDA record
Belmont Instrument LLC
FDA classification date
Apr 25, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential crack in the female connector located on the Heat Exchanger in the set, which will lead to a fluid leak through the crack during priming of the rapid infuser.

Source facts

Action
Belmont Medical notified consignees on about 03/31/2025 via "URGENT MEDICAL DEVICE CORECTION" FedEx or emailed letter. Consignees were instructed to locate affected product and to return the customer Response Form indicating if any affected units were on hand. Consignees were also instructed to follow instructions provided in the letter to inspect affected lots and if any units are identified as having cracks or leaks during priming, they should be discarded. Lot number of the set should be noted and reported to Belmont Medical Technologies. Belmont has indicated that they will replace any product found defective or at the request of the user any affected product that has not been used.
Initiated
2025-03-31
Root cause
Process design
Distribution
Worldwide - US Nationwide distribution in the states of AZ, CA, CO, CT, DC, IL, IN, KS, KY, MA, ME, MN, NC, ND, NJ, NM, OH, OR, PA, SC, TN, TX, VA, VT, WA, WI, WV and the countries of Canada, Hungary, Korea (the Republic of), Latvia, Malaysia.
Recall number
Z-1675-2025
Recall status
Open, Classified
CDRH record ID
213261
FDA classified
2025-04-25
Recall event number
96617

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗