Z-1676-2012: SOLUS Anterior Lumbar Interbody Fusion Spinal Spacer System, Solus Proximal/Distal Deployment Wrenches,

Recall record Z-1676-2012 for Alphatec Spine, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1676-2012
Company in FDA record
Alphatec Spine, Inc.
FDA classification date
May 29, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Alphatec Spine has received several complaints (n+4) on the Alphatec Solus Distal Deployment Wrench for twisting or bending, which has been discovered either during or after instrumentation use. A fifth complaint received in which it was reported that the spline tip of the Alphatec Solus Proximal Deployment Wrench had sheared from the instrument upon use.

Source facts

Action
Alphatec Spine sent an Urgent Medical Device Recall Notification dated March 30, 2012, to all affected customers. The notification letter informed customers of the removal/recall and included instructions on what to do with the recalled products. Initial letter was delivered via FedEx, and followed by telephone call with a total of 4 attempts. Customers were instructed to return all of the affected product. Upon receipt of of the Medical Device Recall Notification, customers were instructed to contact Alphatec Spine's Customer Service Department immediately for instructions on how to return the instruments and fill out the lower portion of this letter to confirm that they have read the notification and have taken all necessary removal actions: Customers should return a signed copy of the form to: Alphatec Spine, Inc. 5818 El Camino Real, Carlsbad, CA 92008 ATTN: Rand Jadan Forms may also be returned via email to rjadan@alphatecspine.com or fax to 760-431-0289. For questions regarding this recall call 760-431-9286.
Initiated
2012-03-19
Root cause
Device Design
Terminated
2012-06-20
Distribution
Worldwide Distribution - USA (nationwide) and Germany
Recall number
Z-1676-2012
Recall status
Terminated
CDRH record ID
108936
FDA classified
2012-05-29
Recall event number
61756

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗