- FDA identifier
- Z-1677-2014
- Company in FDA record
- Ortho-Clinical Diagnostics
- FDA classification date
- Jun 2, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Software Anomaly: Ortho Clinical Diagnostics (OCD) has become aware that if a sample requires follow-up action, such as dilution or retesting due to a mechanical issue, it is possible for that sample to be routed to the sample storage drawer (Storage) prior to the completion of the necessary pending action. A delay in patient test result(s) reporting could occur as operator is not alerted that results have not been produced and that further actions are required.
Source facts
- Action
- US consignees were sent Urgent Product Correction Letters (Cl2014-102, dated 3/20/2014) on 3/20/2014 by FedEx or USPS (for PO Box addresses only) overnight mail, informing them of the issue and instructing them to ensure that all required data elements are entered and in the correct format when using the auto verification feature or custom rules. J&J Foreign affiliate consignees were notified by e-mail on 3/20/2014 informing them of the issue.
- Initiated
- 2014-03-20
- Root cause
- Software design
- Terminated
- 2017-04-11
- Distribution
- Worldwide Distribution-USA (nationwide) including Puerto Rico, and the countries of Australia, Bermuda, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain, and Venezuela.
- Recall number
- Z-1677-2014
- Recall status
- Terminated
- CDRH record ID
- 127236
- FDA classified
- 2014-06-02
- Recall event number
- 68263