Skip to content

Public FDA record

Z-1677-2022: CARTO VIZIGO 8.5F Bi-Directional Guiding Sheath, REF: D138501. Used to introduce various cardiovascular

Recall record Z-1677-2022 for Biosense Webster, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1677-2022
Company in FDA record
Biosense Webster, Inc.
FDA record creation date
Sep 2, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Outside packaging for a device indicated for introducing cardiovascular catheters into the heart, sustained water damage, which may have affected the sterile barrier of the product inside. Devices with a compromised sterile barrier could be a source for infection.

Source facts

Action
On 6/29/22, re call notices started being emailed and mailed to customers who were asked to do the following: 1) The notice should be forwarded to anyone in the facility that needs to be aware. 2) Identify affected devices, unbox devices, and destroy the devices. 3) Complete and return the Business Reply Form and send back the label on empty boxes to receive credit. 4) If any of the affected products have been forwarded to another facility, contact that facility, and provide a copy of this letter to the relevant personnel. 5) Maintain a copy of this communication where the product identified in this letter is located until all products have been destroyed. If you have additional questions about this letter, please email the recalling firm at OneMD-Field-Actions@its.jnj.com
Initiated
2022-06-29
Root cause
Material/Component Contamination
Distribution
US Nationwide distribution in the states of AL, GA, AZ, TX, MA, VA, NY, ME, FL, IA, NM, NH, TN, WI, OH, SC, CA, IN.
Recall number
Z-1677-2022
Recall status
Open, Classified
CDRH record ID
195310
FDA record created
2022-09-02
Recall event number
90647

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗