During SPECT reconstructions using Attenuation Correction and Scatter Correction, no Scatter Correction is being applied in the AutoSPECT Pro application on IntelliSpace Portal. In addition, resolution recovery is not applied correctly in SPECT reconstructions using Astonish in AutoSPECT Pro on Intellispace Portal.
Source facts
Action
On 03/10/2014 the firm sent Urgent Medical Device Correction Letters to their customers.
Initiated
2014-03-11
Root cause
Software design
Terminated
2015-08-04
Distribution
Worldwide distribution, including US nationwide, Argentina, Australia, Austria, Canada, Cayman Islands, Chile, Denmark, Finland, France, Germany, India, Iraq, Israel, Italy, Latvia, Malaysia, Mexico, Netherlands, Panama, Peru, Poland, Saudi Arabia, Singapore, Slovakia, Spain, Switzerland, Turkey and United Kingdom.
Recall number
Z-1678-2014
Recall status
Terminated
CDRH record ID
127106
FDA classified
2014-06-02
Recall event number
68179