Skip to content

Public FDA record

Z-1678-2022: Orthex External Fixation Wire Sharp Bayonet 1.8 x 400 mm Model: K400-18BN

Recall record Z-1678-2022 for OrthoPediatrics Corp. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1678-2022
Company in FDA record
OrthoPediatrics Corp
FDA record creation date
Sep 6, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Manufacturing variance near transition of tip is increasing the likelihood the wire tip may fracture. The fractured tip may be retained in the patient. Retrieval attempts may be made which could result in a delay of surgery and potentially damage soft tissue and increase healing time

Source facts

Action
OrthoPediatrics Corp. provided Urgent Medical Device recall Letter to Distributors on 6/29/22. Letter states reason for recall, health risk and action to take: The affected product must be returned to OrthoPediatrics Corporation. Replacement product will be provided. Please contact customer service to arrange return of the affected product. Distributors should retrieve affected product that may be with customers. No additional corrective actions need to be taken for product already implanted. Please complete and return the attached response acknowledgement form confirming your understanding of these instructions. Any questions regarding this recall can be directed to Joel Batts, Senior VP of Science & Technology, at OrthoPediatrics at jbatts@orthopediatrics.com or 574-367-5396.
Initiated
2022-06-29
Root cause
Process change control
Distribution
US Nationwide Distribution: AZ, AL, CA, FL, FL, MD, MO, NC, NY, TX Foreign: Australia
Recall number
Z-1678-2022
Recall status
Open, Classified
CDRH record ID
195610
FDA record created
2022-09-06
Recall event number
90738

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗