Open original FDA record ↗Recall record
- FDA identifier
- Z-1678-2022
- Company in FDA record
- OrthoPediatrics Corp
- FDA record creation date
- Sep 6, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Manufacturing variance near transition of tip is increasing the likelihood the wire tip may fracture. The fractured tip may be retained in the patient. Retrieval attempts may be made which could result in a delay of surgery and potentially damage soft tissue and increase healing time
Source facts
- Action
- OrthoPediatrics Corp. provided Urgent Medical Device recall Letter to Distributors on 6/29/22. Letter states reason for recall, health risk and action to take: The affected product must be returned to OrthoPediatrics Corporation. Replacement product will be provided. Please contact customer service to arrange return of the affected product. Distributors should retrieve affected product that may be with customers. No additional corrective actions need to be taken for product already implanted. Please complete and return the attached response acknowledgement form confirming your understanding of these instructions. Any questions regarding this recall can be directed to Joel Batts, Senior VP of Science & Technology, at OrthoPediatrics at jbatts@orthopediatrics.com or 574-367-5396.
- Initiated
- 2022-06-29
- Root cause
- Process change control
- Distribution
- US Nationwide Distribution: AZ, AL, CA, FL, FL, MD, MO, NC, NY, TX Foreign: Australia
- Recall number
- Z-1678-2022
- Recall status
- Open, Classified
- CDRH record ID
- 195610
- FDA record created
- 2022-09-06
- Recall event number
- 90738