Z-1679-2014: 115 V Blanketrol II, Model 222S Hyper-Hypothermia System . Used to lower or to raise a patients temperat

Recall record Z-1679-2014 for Cincinnati Sub-Zero Products Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1679-2014
Company in FDA record
Cincinnati Sub-Zero Products Inc
FDA classification date
Jun 10, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The incorrect voltage component was placed into the device during manufacturing. This may make the device unable to maintain patient temperature during cooling mode. The device will continue to indicate that the unit is cooling because the compressor will continue to run, however water in the reservoir will begin to return to ambient temperature. The actual water temperature will be displayed but

Source facts

Action
The firm initiated a telephone notification on 01/08/2014, followed by a written notification which was sent to the customers via letter or email on 01/31/2014. URGENT Medical Device Field Action (January 31, 2014) REF: 115V Blanketrol II Model 22S Hyper-Hypothermia System (1) Discontinue Use and contact CSZ (2)Upon receipt of new 115V (p/n 91113) solenoid coil(s), immediately remove and destroy/discard all the affected 220V (p/n 3600) solenoid coil(s) (3)After the affected 220V (p/n 36007) solenoid coil(s) have been removed, discard and replaced, please complete and return the enclosed response form as soon as possible to acknowledge receipt of this notification and to imform CSZ that you have performed and completed the requested actions. Return the form by fax
Initiated
2014-01-08
Root cause
Mixed-up of materials/components
Terminated
2014-06-06
Distribution
The product was distributed to one customer in MI, and internationally (Taiwan)
Recall number
Z-1679-2014
Recall status
Terminated
CDRH record ID
127458
FDA classified
2014-06-10
Recall event number
68356

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗