Z-1679-2020: ONCOR Avante Garde #5863472, ONCOR Expression #7360717, ONCOR Impression #5857920, ONCOR Impression Plus

Recall record Z-1679-2020 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1679-2020
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA record creation date
Apr 7, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential safety issue with the dose monitoring system safety interlocks which could result in a 10-20% overdose of radiation therapy.

Source facts

Action
A Customer Safety Advisory Notice was sent to all affected consignees on 2/18/2020 via Adobe Mega Sign or USPS certified mail. The notice describes the interlock scenarios, provides user instructions, and warns the user of potential risks. The interlock scenarios are as follows: MONITOR 2: Monitor 2 total dose coincidence. Intent: This interlock is intended to prevent overdose when the primary dose monitory system fails to complete the treatment. TREATMENT TIME: Preset treatment time coincidence with actual treatment time. Intent: This interlock is intended to prevent overdose if preset treatment time is exceeded. MONITOR SYNC (SW): Excessive accumulated dose deviation detected between the primary and the secondary dose monitoring systems. Intent: This interlock is intended to prevent patient mistreatment due to the inaccurate monitoring of delivered dose. When one or more of the above described interlocks occur, it may be assumed that there is an error in the dose monitor system.The Consumer Safety Advisory Notice states that the Operator's Manual Instructions Update document should be added to the product's manual and all users should be informed of the issues.
Initiated
2020-02-18
Root cause
Use error
Terminated
2021-02-09
Distribution
Worldwide distribution. US consignees in the following states: Alabama, California, Delaware, Florida, Georgia, Iowa, Illinois, Kentucky, Louisiana, Massachusetts, Maryland, Maine, Michigan, Minnesota, Missouri, Mississippi, North Carolina, Nebraska, New Jersey, Nevada, New York, Ohio, Oklahoma, Pennsylvania, Puerto Rico, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Wisconsin, West Virginia, Wyoming
Recall number
Z-1679-2020
Recall status
Terminated
CDRH record ID
180603
FDA record created
2020-04-07
Recall event number
85161

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗