Skip to content

Public FDA record

Z-1679-2022: Abbott Alinity m SARS-CoV-2 CTRL Kit, REF 09N78-085, each kit contains 12 x 1.3mL CTRL - and 12 x 1.3 mL

Recall record Z-1679-2022 for Abbott Molecular, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1679-2022
Company in FDA record
Abbott Molecular, Inc.
FDA record creation date
Sep 6, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Reports received of an increase of message code 9186 (internal control failed) when using the kit.

Source facts

Action
The firm issued recall letters dated 8/19/2022 via FedEx priority overnight express on 8/19/2022. The letters provided the background, potential impact, and necessary actions. The necessary actions included contacting Abbott Customer Support for replacement of any unused kits in the consignee's inventory of the affected lot number and discard any inventory in their possession. If replacement inventory is not available, the consignee was told they may continue to use the affected product until replacement material is obtained. They are to ensure that valid Control values are obtained for all control levels prior to reporting results as specified in the associated package Insert. A Customer Reply Form was enclosed for completion and return. If the consignee has forwarded any kits of the affected lot to other laboratories, they are instructed to inform them of the recall, provide them a copy of the letter, and have them take the necessary actions listed. The letter also instructs the consignee to inform their laboratory personnel of the issue and retain a copy of the communication for further reference.
Initiated
2022-08-19
Root cause
Under Investigation by firm
Distribution
Distribution was made to AZ, CA, CT, FL, GA, IL, KS, KY, LA, MA, MI, MN, MO, NC, NJ, NV, NY, OK, PA, RI, TN, TX, WA, and WI. There was government distribution but no military distribution. Foreign distribution was made to Chile and Taiwan.
Recall number
Z-1679-2022
Recall status
Open, Classified
CDRH record ID
195729
FDA record created
2022-09-06
Recall event number
90789

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗