Open original FDA record ↗Recall record
- FDA identifier
- Z-1680-2013
- Company in FDA record
- Linvatec Corp. dba ConMed Linvatec
- FDA classification date
- Jul 10, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Conmed Linvatec received complaints that their products C7120 APEX Arthroscopy Tubing Set, One Connection and C7122 APEX Arthroscopy Tubing Set, Two Connection, contained a scratch/grooved area on the inside diameter of the pressure fitting that may cause the APEX Irrigation (pump) Console to sense a pressure drop and increase fluid flow to the surgical site.
Source facts
- Action
- ConMed Linvatec sent an "URGENT: Medical Device Recall Notification" letter by FedEx next day delivery or equivalent on 5/30/2013 to all domestic direct consignees. ConMed Linvatec is translating customer notifications letters into the necessary languages and will send International notification letters no later than 6/4/2013.
- Initiated
- 2013-05-30
- Root cause
- Equipment maintenance
- Terminated
- 2013-09-04
- Distribution
- Worldwide Distribution-USA (nationwide) including the states of AR, CA, FL, GA, IL, IN, KS, MO, NC, NJ, NM, NY, OH, OK, PA, TN, TX, UT, VA, and WI, and the countries of Austria; Belgium, Canada, China, Czech Republic, Finland, France, Great Britain, Italy, and Japan.
- Recall number
- Z-1680-2013
- Recall status
- Terminated
- CDRH record ID
- 118550
- FDA classified
- 2013-07-10
- Recall event number
- 65320