Skip to content

Public FDA record

Z-1680-2022: Optima Coil System Model OPTI0520CSF10. For endovascular embolization.

Recall record Z-1680-2022 for BALT USA, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1680-2022
Company in FDA record
BALT USA, LLC
FDA record creation date
Sep 6, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Products were released to customers prior to the approval and release of a test report authorizing the use of an existing material from an alternate supplier.

Source facts

Action
On or about 07/15/2022, Balt USA LLC communicated an "URGENT FIELD SAFETY NOTICE (FSN)/PRODUCT RECALL" Letter informing customers that Balt became aware that finished goods were released and distributed to customers prior to the approval and release of a test report authorizing the use of existing material from an alternate supplier. On 08/08/2022 and updated/revised "URGENT MEDICAL DEVICE RECALL" letter was sent via certified mail to customers re-emphasizing the recall and providing updated instructions Customer are instructed to: -Inform within their hospital safety officers, pharmacists, head of neuroradiology and the neuroradiology department staff, as well as any other person. -Return to Balt USA any unused product from lot #F220600470: +Collect and put in quarantine the Optima products concerned by this recall and then return them to Balt USA through the usual "RMA" (Return Material Authorization) procedure by contacting Balt USA's Customer Service Department +Keepj informed Balt USA about the status of every unit of Optima products concerned by this recall; +Fulfill the "Notice of receipt" then return to Balt USA via the indicated contact. -If product from lot #F220600470 has been used, no further action is required. For question contact - Quality Department at QA@baltgoup.com or telephone number 949-788-1443 or Fax Number: 949-788-1444
Initiated
2022-07-26
Root cause
Under Investigation by firm
Distribution
U.S.: AZ and FL O.U.S.: None
Recall number
Z-1680-2022
Recall status
Open, Classified
CDRH record ID
195570
FDA record created
2022-09-06
Recall event number
90724

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗