Skip to content

Public FDA record

Z-1681-2020: Thermage Face Tip 3.0, REF/UDI: TTNS3,00C3-400/UDI (01)000850608002377; REF TTNS3.00C3-600/UDI (01)000850

Recall record Z-1681-2020 for Solta Medical Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1681-2020
Company in FDA record
Solta Medical Inc
FDA record creation date
Apr 7, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Complaints of sparking of the tips of the electrosurgical device used in RF surgical procedures. The use of the affected treatment tips could contribute to potential skin burns.

Source facts

Action
On 12/30/2019, the firm sent an "URGENT VOLUNTARY MEDICAL DEVICE RECALL" to customers via registered mail to inform them that they have become aware that there is a potential problem related to their device that may result in patient burns. This recall affects 20 lots manufactured and distributed between 08/14/2019 and 12/12/2019. The firm is instructing the customers to: 1. Review their inventory for all devices impacted by this voluntary recall and remove them from the inventory. 2. The manufacturer/ recalling firm is asking their customers to contact Customer Service at 1-877-782-2286 option 1 to arrange the return of the impacted devices in the inventory. 3. In addition, the firm is requesting the customers to complete the enclosed Product Recall Acknowledgement Form and return it to the manufacturer via email at orders@solta.com or fax at 1-800-685-6430 within 5 days. Solta Customer Service has followed up with customers via phone call and/or email.
Initiated
2019-12-30
Root cause
Process design
Terminated
2024-04-29
Distribution
US - AK, AR, AZ, CA, FL, GA, IL, NY, PA, TX. UT, and WA OUS -Japan, Korea, Malaysia, Singapore, Taiwan, and Thailand.
Recall number
Z-1681-2020
Recall status
Terminated
CDRH record ID
180495
FDA record created
2020-04-07
Recall event number
85314

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗